3-2 | Model 888 ATEX IECEx Tail Gas Analyzer
Analyzer Site Preparation
The 888 Tail Gas Analyzer is approved for installation in hazardous areas
classified as Zone 1, ( Per ATEX & IECEX).
The analyzer must be installed on a horizontal run of the tail gas pipe. A
site plan should be prepared defining the system mounting location and
provisions for power/data interconnect wiring. It is highly recommended
that a servicing platform equipped with safety railings be installed at the
analyzer mounting site.
Unpacking and Inspection
Remove components from the packing case(s) carefully; check contents
against packing list. Open the enclosures and inspect all components for
obvious damage and broken/loose parts or fittings. The Demister/Inlet
Valve assembly is packaged separatly and shipped within the analyzer
crate to avoid damage during transportation.
Notify the carrier and AMETEK Service (1-800-222-6789) immediately if
parts are missing or damage is found.
Space Requirements
Refer to Figure 3-1 (or the “Field Unit Installation” drawing from the Cus-
tomer Drawing Package). The analyzer weighs approximately 105 kg (230
lbs).