Installation and Start-Up | 3-35
From the Temperature Control dialog box (Figure 3-15), verify the
Setpoint, Too Hot, and Too Cold setpoint temperatures for each
zone are correct. See also “Temperature Control” in Chapter 4.
If you need to alter the temperature setpoints or ranges, refer to
Temperature Control” in Chapter 4 before making changes.
After configuring the temperature zone setpoints, click OK then
Apply. Save the changes to EEPROM.
g. Check the analyzer temperature zones:
While working from the Configurator Software (Analyzer Data
tab), view and record the current (start-up) temperature readings
for each zone. These values can be used as a reference later while
comparing the temperature values after the analyzer has warmed
up. This will also help you determine if the signal wiring is prop-
erly terminated and if all of the circuitry is operating properly.
NOTE
Status (tab)Temp
Control
Figure 3-14.
Analyzer Data tab
(Model 932S).