Installation and Start-Up | 3-37
Below each configured channel (Assignment), confirm that the
Species Iout/Vout Low Scale and Iout/Vout High Scale are the cor-
rect minimum and maximum ranges required.
To ensure the analog output for each channel does not go beyond
the low and the high scales (i.e., limited to 4 mA and 20 mA), en-
able (check) the Limit to 4..20 mA check box.
If you need to alter the ranges, refer to “Output Setup” in Chapter 4
and “Analog Output Calibration” in Chapter 5 before making
changes.
After configuring the Current and Voltage Output Channel ranges,
click OK then Apply. Save the changes to EEPROM.
i. From the Alarm Setting and Stream Switch Parameters dialog
box, verify that the Concentration Alarms (1 and 2) are configured
correctly for the application.
If you need to alter the ranges, refer to “Alarm Setting and Stream
Switch Parameters” in Chapter 4.
h. From the Output Setup dialog box, verify the current and volt-
age output ranges are configured correctly for the application.
Outputs will vary, depending on the application and analyzer.
Model 931S Analyzers are typically equipped with only one V/I
output, while Model 932S Analyzers may have up to four.
NOTE
NOTE
Status (tab)
Output Setup
Status (tab)Alarm/
Stream Switch
Figure 3-17.
Output Setup dialog box.