Calibration SG Series Programming
7-14 M550129-03 Rev L
7.8.1 Adjustment for Current Mode
STEP DESCRIPTION
1. Set the SGI Series unit to operate in remote current programming mode using
an external 0-5 VDC voltage source as shown in Figure 7-3.
+
-
1
13
23 IP RTN
COM 6
+
-
16 IP 10V
IP 5V 10
0-5 VDC
VOLTAGE
SOURCE
0-10 VDC
VOLTAGE
SOURCE
Figure 7-3. Remote Current Programming Using 0-5 VDC or 0-10 VDC Voltage Source
2. Attach a precision meter across the shunt Kelvin terminals.
3. Attach a precision meter in parallel with the voltage programming source.
4. Set the voltage source to 0.0V.
5. Apply AC power and press the "Output On/Off" switch (Figure 7-4) to the ON
position.
Figure 7-4. SGI Front Panel
6. Set the voltage source to 5.0V ± 1mV.
On/Off Switch