Ethernet Configuration and Remote Programming SG Series Programming
5-10 M550129-03 Rev L
Table 5-2. Remote Mode Power-on Conditions
Condition
Default
Voltage
0 Volts (initial from factory poweron voltage); otherwise, last
value saved by SCPI command or by the SAVE SETTINGS
button in the Web Settings page.
See CAL:INIT:VOLT to change.
Current
0 Amps (initial from factory poweron current); otherwise, last
value saved by SCPI command or by the SAVE SETTINGS
button in the Web Settings page.
See CAL:INIT:CURR to change.
Soft Voltage Limit
Model maximum voltage *
Soft Current Limit
Model maximum current *
OVP Trip Voltage
Model maximum voltage +10% (initial from factory poweron
OVP); otherwise, last value saved by SCPI command or by
the SAVE SETTINGS button in the Web Settings page.
See CAL:INIT:VOLT:PROT to change.
Delay
0.5 seconds
Foldback Protection
OFF (non-configurable)
Output
ON ** See CAL:MOD:POWERON
Hold
OFF
Unmask
NONE
Service Request Capability
OFF
* User-programmable temporary limit (reverts to power-on defaults after power
cycle or Reset command is issued).
** User-selectable
5.3 EXTERNAL USER CONTROL SIGNAL
CONNECTOR
A10-pin Molex connector (Figure 5-7) located at the rear panel provides
external auxiliary control signals to increase the user’s operating control of
the supply. The mating receptacle is Molex 43025-1000 with 10 female
terminals. The Molex terminals accommodate wire sizes from #20 - #24.
The relay outputs, when active, connect the POLARITY, ISOLATION and
SENSE pins (Pins 6, 7 and 8) of the connector to the relay COMMON pin
(Pin 5). The relays are rated at 120VAC/125VDC @ 1A. Any change in
output (voltage, current, etc.) initiated by the user from the RS232, GPIB, or
Ethernet interface, will generate a 10ms synchronization pulse at the rear
panel User Control Signal Connector of the unit (TRIGGER OUT).