SG Series Programming IEEE/RS232 Configuration and Remote Programming
M550129-03 Rev L 3-9
3.4 REMOTE PROGRAMMING VIA RS232
The RS232 interface operates at fixed 19.2K baud for SGA/SGe and is
selectable from 2400 to 19.2K baud for the SGI, with 8 data bits, no parity,
and 1 stop bit.
All commands are supported at the RS232 interface with the exception of
the Service Request (SRQ) function, which is a GPIB-specific function
requiring the dedicated Service Request line of the IEEE 488.2 interface. In
this case, the SRQ function has no effect. The RS232 interface is accessible
through the rear panel RJ-type 6P6C (6-pin) connector (Figure 3-8), labeled
RS232 on the power supply’s rear panel (see Figure 3-1, Figure 3-2 and
Figure 3-3).
Figure 3-8. RS232 Rear Panel RJ-type 6P6C Connector Pinout