SCPI Command Operation SG Series Programming
6-24 M550129-03 Rev L
Command
Description
SCPI
:TERM
<NRf>
Sets the return string terminators to be used by the device. Factory
set to 3. The valid range is 1-4. Values indicate the following
terminator(s):
1 = 0x0d only (CR), 2 = 0x0a only (LF), 3 = 0x0d 0x0a (CR LF), 4 =
0x0a 0x0d (LF CR)
N
:TERM?
Returns the string terminators to be used by the device.
N
:VERsion?
Returns a numeric value corresponding to the SCPI version number
for which the instrument complies. The response is in the format
YYYY.V where the Y’s represent the year and V represents the
approved version number for that year (e.g., 1995.0)
C
Table 6-6 Table 6-7. System Fault Registers
Bit Position
Bit Weight
Fault18
Fault916
Fault1724
Fault2531
7
128
Channel 1
not used
not used
not used
6
64
not used
not used
not used
not used
5
32
not used
not used
not used
not used
4
16
not used
not used
not used
not used
3
8
not used
not used
not used
not used
2
4
not used
not used
not used
not used
1
2
not used
not used
not used
not used
0
1
not used
not used
not used
not used
The SYStem:FAULt? query returns 4 numeric values separated by commas.
Each value is the decimal equivalent of the total bit weights for that System Fault
Register as described in the table above.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Ametek Sorensen SG Series and is the answer not in the manual?

Ametek Sorensen SG Series Specifications

General IconGeneral
BrandAmetek
ModelSorensen SG Series
CategoryPower Supply
LanguageEnglish

Related product manuals