SCPI Command Operation SG Series Programming
6-30 M550129-03 Rev L
Returns the value of the gain for current
measurements.
Sets the value of the offset for current
measurements.
Returns the value of the offset for current
measurements.
Sets the current measurement calibration point
(1 or 2). The actual output current is measured with
an external meter.
Calibrate Measure Voltage subsystem
Returns the integer value of the A/D for the voltage
measurement.
Calculates the value of the gain and offset for voltage
measurements.
Sets the value of the gain for voltage measurements.
Returns the value of the gain for voltage
measurements.
Sets the value of the offset for voltage
measurements.
Returns the value of the offset for the voltage
measurements.
Sets the voltage measurement calibration point
(1 or 2). The actual output voltage is measured with
an external meter.
Calibrate Model subsystem
(Ethernet only) Sets the date last calibrated;
format: MM DD YYYY (space after MM and DD
required)
(Ethernet only) Returns the date last calibrated.
(Ethernet only) Sets the date next calibration is
required;
format: MM DD YYYY (space after MM and DD
required)
(Ethernet only) Returns the date next calibration is
required.
Sets the default output enable condition at power on.
Input format:
“ON,INIT” enables the output at next power on
“OFF,INIT” disables the output at next power on
NOTE: Quotation marks are required in the
command string.
Returns the status of the output enable condition at
power on
Sets the default output enable condition when the
*RST command is issued.
Input format:
“ON,INIT” enables the default output
“OFF,INIT” disables the default output
NOTE: Quotation marks are required in the
command string.
Returns the status of the default output enable
condition when the *RST command is issued