E-2 | Thermox Series 2000 WDG-IV / IVC / IVCM
PARAMETER DEFAULT SETTINGS CUSTOMER SETTINGS
SENSOR CONFIGURE O2
CEM/O2, INSITU
WDG-IV, WDG-HPII
SENSOR CONFIGURE COMB WDG-IVC, WDG HPIIC
SENSOR CONFIGURE METHANE WDG-IVCM
SENSOR NODE ADDRESS 0
BAUD RATE 9600
ENABLE SERIAL ENABLED
COMBUSTIBLE TYPE HOTWIRE
COMBUSTIBLE TYPE RTD
COMBUSTIBLE SCALE PPM
PROCESS PRESSURE 0.00 PSIG
O2 SPAN GAS SETPOINT 20.90%
O2 ZERO GAS SETPOINT 2%
COMB SPAN GAS SETPOINT 1600 PPM
METHANE GAS SETPOINT 2.00%
AUTO CAL CYCLE TIME 8:00
AUTO CAL CYCLE FREQ 1 DAY
AUTO VERIFY CYCLE TIME 9:00
AUTO VERIFY CYCLE FREQ 1 DAY
AUTO CAL/VERIFY DAYS/HRS DAYS,DAYS
AUTO CAL ENABLE/DISABLE ENABLED
AUTO VER ENABLE/DISABLE ENABLED
O2 SPAN GAS DURATION 2:00 MINUTES
O2 ZERO GAS DURATION 2:00 MINUTES
COMB SPAN GAS DURATION 2:00 MINUTES
RECOVERY DURATION 4:00 MINUTES
K FACTOR 1
OUTPUT # 1 FUNCTION % OXYGEN
CURRENT MODE,RANGE 4 TO 20 MA = 0-10
TRACK/HOLD CAL VERIFY HOLD,TRACK
OUTPUT FILTER # 1 100