EasyManua.ls Logo

Anritsu MS2665C - Page 172

Anritsu MS2665C
279 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SECTION 6 PERFORMANCE TESTS
6-41
Step Procedure
9 Change the attenuator in 10 dB steps, set the reference level as shown in the table below and
read the marker level each time.
Reference level
setting
0 dBm
–10 dBm
–20 dBm
–30 dBm
–40 dBm
–50 dBm
–60 dBm
–70 dBm
–80 dBm
dBm
dBm
dBm
dBm
dBm
dBm
dBm
dBm
dBm
dB
dB
dB
dB
dB
dB
dB
dB
dB
dB
dB
dB
dB
dB
dB
dB
dB
dB
Error
Correction factor
of ATT
Marker
readout
10 Calculate the error from the following equation.
Error = Marker readout - reference level set value - correction factor of ATT