EasyManuals Logo

Applied Biosystems 7500 Getting Started Guide

Applied Biosystems 7500
106 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #66 background imageLoading...
Page #66 background image
Chapter 5 Generating Data from RQ Plates – 7500 Fast System
Troubleshooting
56 Applied Biosystems 7300/7500/7500 Fast Real-Time PCR System Relative Quantification Getting Started Guide
Notes
FAST
FAST
FAST
Troubleshooting
Troubleshooting
Observation Possible Cause Action
High C
T
values/poor
precision or failed PCR
reactions
Target is difficult to amplify Increase the annealing/extension time in the thermal
cycler protocol.
Increase the annealing/extension temperature to 62 ° C.
Insufficient cDNA template is
present
Use 10 to 100 ng of cDNA template per 20-µL reaction.
Quality of cDNA template is poor 1. Quantify the amount of cDNA template.
2. Test the cDNA template for the presence of PCR
inhibitors.
Sample degradation Prepare fresh cDNA, then repeat the experiment.
The TaqMan 2 Universal PCR
Master Mix was used instead of
the TaqMan Fast Universal PCR
Master Mix (2)
, No AmpErase
UNG
Prepare the reactions with the correct Master Mix.
Primer-dimer formation To ensure optimal results, run the reaction plate as soon as
possible after completing the reaction setup. If you cannot
run a reaction plate within 2 hours after completing the
reaction setup, refrigerate or freeze the reaction plate until
you can load and run it on the 7500 Fast instrument.
Low R
n
or R
n
values Extension time is too short Use the default thermal profile settings (see page 49).
Primer-dimer formation To ensure optimal results, run the reaction plate as soon as
possible after completing the reaction setup. If you cannot
run a reaction plate within 2 hours after completing the
reaction setup, refrigerate or freeze the reaction plate until
you can load and run it on the 7500 Fast instrument.
Run takes more than 40
minutes
Thermal cycler mode is set to
Standard or 9600 Emulation
Make sure that the thermal cycler mode is set to Fast (see
page 50).
Rn vs. Cycle plot is not
displayed
ROX dye was not selected as the
passive reference when the plate
document was set up
Select ROX dye as the passive reference when you set up
the plate document.
Extremely high R
n
or
R
n
values
ROX dye was not selected as the
passive reference when the plate
document was set up
Select ROX dye as the passive reference when you set up
the plate document.
Evaporation Make sure that the reaction plate is sealed completely,
especially around the edges.

Table of Contents

Other manuals for Applied Biosystems 7500

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Applied Biosystems 7500 and is the answer not in the manual?

Applied Biosystems 7500 Specifications

General IconGeneral
Detection MethodFluorescence
Sample Capacity96-well plate
Thermal CyclerYes
Temperature Uniformity±0.5°C
TypeReal-Time PCR System
Thermal Accuracy±0.25°C
Detection Channels5
Weight45 kg
Power Requirements100-240 V, 50/60 Hz
Dynamic Range9 logs
Reaction Volume Range10-100 µL
Temperature Range4°C to 100°C
Thermal Cycler Block96-well
SoftwareSDS Software
DetectionMultiplex (up to 5 colors)

Related product manuals