Chapter 6 Analyzing Data in an RQ Study
Configuring Analysis Settings
Applied Biosystems 7300/7500/7500 Fast Real-Time PCR System Relative Quantification Getting Started Guide 63
Notes
6
Configuring Analysis Settings
After you create the RQ Study document, you must
specify parameter values for the analysis.
Unless you have already determined the optimal
baseline and threshold settings for your experiment,
use the automatic baseline and threshold feature of the
SDS Software 1.3.1 (auto Ct), explained below. If the
baseline and threshold were called correctly for each
well, you can proceed to view the results. Otherwise,
you must manually set the baseline and threshold as
explained in “Manual Baseline and Threshold
Determination” on page 65.
To configure analysis settings:
1. Click or select Analysis > Analysis Settings.
2. In the Detector drop-down list, select All.
3. Select Auto Ct. The SDS Software 1.3.1
automatically generates baseline and threshold
values for each well.
IMPORTANT! After analysis, you must verify
that the baseline and threshold were called
correctly for each well, as explained in
“Adjusting the Baseline and Threshold” on
page 65.
Alternatively, you can select Manual Ct and
specify the threshold and baseline manually.
4. Select the Calibrator Sample.
Note: If your experiment uses only a single
plate, there must be at least two different samples
that have different names and have their own
endogenous controls. (You can go back to a
saved RQ Plate document and change the sample
names, if necessary.)
5. Select the Endogenous Control Detector.
2
3
4
5
6
7
8

Table of Contents

Other manuals for Applied Biosystems 7500

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Applied Biosystems 7500 and is the answer not in the manual?

Applied Biosystems 7500 Specifications

General IconGeneral
Detection MethodFluorescence
Sample Capacity96-well plate
Thermal CyclerYes
Temperature Uniformity±0.5°C
TypeReal-Time PCR System
Thermal Accuracy±0.25°C
Detection Channels5
Weight45 kg
Power Requirements100-240 V, 50/60 Hz
Dynamic Range9 logs
Reaction Volume Range10-100 µL
Temperature Range4°C to 100°C
Thermal Cycler Block96-well
SoftwareSDS Software
DetectionMultiplex (up to 5 colors)

Related product manuals