Calibration of the thermostat mixer P300
1 Remove the calibration label attached to the control
panel. (See fig. 1)
2 With power to panel and no water present in the tub,
the display shows “--”. (See fig. 2)
3 Press the hidden button and up button
simultaneously (See fig. 3) until the buzzer sounds and
the display shows “code”.
WARNING
Failure to calibrate the thermostat mixer may cause scalding.
WARNING
Always use a calibrated thermometer to check water temperatures against the panel
display valve.
NOTE
To exit the “code” screen at any point press the hidden button and down button
simultaneously until “--” is displayed.
NOTE
Anytime water is in contact with the bathtub temperature sensor, the display will give
a temperature value.
WARNING!
Do not use before calibration.