PRODUCT INFORMATION Maxi Move
Total weight (standard jib) 63.5 kg (140 lb)
Lifting capacity
Regular jib: 227 kg (500 lb)
Extended jib: 130 kg (287 lb)
Battery pack weight 3.8 kg (8.4 lb)
Turning radius 1222 mm (48 in)
Minimum door requirement 717 mm (28.25 in)
Operating forces of controls 2.5- 3N
ELECTRICAL
Degree of protection
IPx7 - Hand control
IPx4 - Maxi Move
Internally powered 24 Vdc
Control voltage output 24 Vdc
Battery charger (part #700.24250) input 100 to 240 Vac
Up and down current limiting 12 ± 1 Amp.
Duty cycle 15% - max. 2 min. continuous use
Sound power level up 61.9 dBA
Sound power level down 61.7 dBA
Medical equipment Type B protection against electrical shock in accordance with IEC 60601-1
ARJO products meet the requirements of Electromagnetic Compatibility (EMC) as stated in clause 12.5 of the Medical Devices Directive 93/42/EEC.
The Maxi Move conforms to UL STD 2601-1, CSA C22.2 No. 60601.1 - M90 (except when being used in conjunction with a stretcher; see “Using your Maxi
Move” section on page 24), and ISO 10535:2006.
DIGITAL SCALE SPECIFICATIONS
Weight range 227 kg (500 lb)
Display resolution and type 0.1 kg (0.2 lb), liquid crystal display
2-50 kg ±100 g 4-100 lb ±0.2 lb
Accuracy (in service) Class III 50-200 kg ±200 g 100-400 lb ±0.4 lb
200-227 kg ±300 g 400-500 lb ±0.6 lb
OPERATION AND STORAGE CONDITIONS
Ambient temperature range (lift) Operation: 10° to 40°C (+50 to +104 F) Storage: - 40° to 70°C (-40 to 158F)
Ambient temperature range (batteries) Operation: 10° to 40°C (+50 to +104 F) Storage: -15° to 40°C (+5 to +104 F)
Relative humidity range Operation: 30 to 75% Storage: 10 to 80%, incl. condensation
Atmospheric pressure range Operation: 700 hPa to 1060 hPa Storage: 500 hPa to 1060 hPa
RECYCLING
Battery
Sealed lead-acid, rechargeable,
recycleable
Package Cardboard recycleable
The Lift Separated and recycled, according to the European Directive 2002/96/EG (WEEE).
WARNING: Radio transmitting devices such as mobile telephones, two-way radios,
etc., should never be used near the Maxi Move, since they can interfere with the
function of the lift. Cables from potentially strong sources of electromagnetic fields
should not be placed near the unit.
WARNING: Equipment not suitable in the presence of flammable anesthetic
mixture with air or oxygen, or with nitrous oxide.