Chapter 3 – Start up, adjust and measures
UM-23900B-U User manual ATEQ CDF60 Page 15/40
3.3. STANDARD CONDITIONS (STD COND)
This parameter allows displaying the measurements values corrected to the standards
conditions values, atmospheric pressure and temperature, to the user preferences.
Enter into the parameters menu, by pressing and
old the ENTER
during 3 seconds.
+/- 40.00 cm3/min
AIR
FLOW : 09.36
TEMP. : 25.8°C
ATM.P : 1010.5 hPa
The PARAMETER menu is displayed. By using the
UP
and DOWN key select
STAND. COND parameter.
FILTER : 1.0 s
FLOW UNIT.: cm3/min
.:
Then press the ENTER key to modify the
parameter (the cursor moves on the right).
FILTER : 1.0 s
FLOW UNIT.: cm3/min
.:
By using the UP and DOWN key
validate the function by YES. Validate with the
ENTER
key.
FILTER : 1.0 s
FLOW UNIT.: cm3/min
.:
The standards temperature and atmospheric
pressure values are displayed, it's possible to
modify following the user preferences. Default
values: 20° C and 1013.2 hPa.
>TEMP. : 20.0°C
PRESS. : 1013.2hPa
To return to the measurement mode, press twice
CANCEL
. The measurements values
displayed are corrected following standards
conditions.
.
cm
m
n
AIR STD.COND.
FLOW : 09.36
TEMP. : 25.8°C
ATM.P : 1010.5 hPa