Chapter 3 – Start up, adjust and measures
UM-23900B-U User manual ATEQ CDF60 Page 19/40
3.7. AUTOMATIC POWER OFF (AUTO OFF)
This feature conserve battery life by automatically turning the instrument off after a
user-defined period of time within there has been no key activity.
Enter into the parameters menu, by pressing and
old the ENTER
during 3 seconds.
-
40
.
00
cm
3/
m
i
n
AIR STD.COND.
FLOW : 09.36
TEMP. : 25.8°C
ATM.P : 1010.5 hPa
The PARAMETER menu is displayed. By using the
UP
and DOWN key select
AUTO OFF parameter.
PARAMETER
GAZ : AIR
TEMP.UNIT.: CELSIUS
>
AUTO OFF:
0
5 min
Then press the ENTER key to modify the
parameter (the cursor moves on the right).
PARAMETER
GAZ : AIR
TEMP.UNIT.: CELSIUS
A
UTO OFF:
0
5 min
<
By using the UP and DOWN key to
modify the switch off delay value.
PARAMETER
GAZ : AIR
TEMP.UNIT.: CELSIUS
A
UTO OFF:
0
7 min
<
Validate with the ENTER key.
PARAMETER
GAZ : AIR
TEMP.UNIT.: CELSIUS
>
A
UTO OFF:
0
7 min
To return to the measurement mode, press
CANCEL
.
-
40
.
00
cm
3/
m
i
n
AIR STD.COND.
FLOW : 09.36
TEMP. : 25.8°C
ATM.P : 1010.5 hPa
It's possible to disable the AUTO OFF function and
leave continuously the instrument on, for that by
using the DOWN
key go the minimum
(inferior to 1 minute) then appears "DISABLE"
Validate with the
ENTER
key. The
instrument will be power on continuously.
PARAMETER
GAZ : AIR
TEMP.UNIT.: CELSIUS
>
A
UTO OFF: DISABLE