BarcoMed control panel settings Single Source Master
68
applications, which can cause image color values to appear
distorted in the background application.
As shown in figure 3, the 20 standard system colors are
converted from RGB to gray values. The rest of the 236 entries
from index 10 to 245 contain the missing gray values so that the
palette has the full 256 gray values within it.
Figure 3: Static Gray Palette with system colors
Note: If you are unsure whether or not your application requires
this “Static Gray Palette with System colors” mode, contact your
application provider.
Static Gray Palette with NO system colors
This option sets the palette to be a static linear ramp of 256
shades of gray. Therefore, applications are denied the ability to
dynamically change or allocate palette entries. This prevents
palette conflicts between applications, which can cause image
color values to appear distorted in the background application.
As shown in figure 4, each of the 256 entries in the palette has
an RGB value of (i, i, i) where i is the index from 0 to 255.
Figure 4: Static Gray Palette with NO system colors
If you wish to use a static gray palette we recommend using the
“Static Gray Palette including Standard System Colors” option
instead of this one. This is due to the fact that some applications
assume that the first and last 10 entries of the palette are the