Manually Controlling the Biomek NX in Biomek Software
Using Advanced Manual Control
4
2
Choose Advanced Controls.
3
Select Pod1. Advanced Manual Control for the pod appears (Figure 4.5). The current position is
shown in Current Position.
4
Choose Close to close Advanced Manual Control.
5
Choose Exit to close Manual Control.
Moving the Pod to a Safe Roving Height
Use Move Z-Max in Advanced Manual Control to move the pod to its highest configured height. This
helps to avoid collisions when moving the pod around the deck manually.
NOTE The possibility of collisions is not completely eliminated by this command; for example, if the gripper
is extended and holding a microplate, and the pod is moved over a tip box on a tiploader, a collision could
occur.
1
Choose Instrument > Manual Control. Manual Control appears (Figure 4.2).
2
Choose Advanced Controls.
3
Select Pod1. Advanced Manual Control for the pod appears (Figure 4.5).
4
Choose Move Z-Max.
5
Choose Close to close Advanced Manual Control.
6
Choose Exit to close Manual Control.

Table of Contents

Other manuals for Beckman Coulter Biomek NXP

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Beckman Coulter Biomek NXP and is the answer not in the manual?

Beckman Coulter Biomek NXP Specifications

General IconGeneral
BrandBeckman Coulter
ModelBiomek NXP
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals