5.3 Configuration
Operation
5290 / 06/2023 enBENNING PV 1-1+26
Procedure for PVsafety measuring lines
1. Connect the black PVsafety measuring line to the black PVtest socket of the device.
2. Connect the red PVsafety measuring line to the red PVtest socket of the device.
5.3 Configuration
5.3.1 Setting the “Auto-Power-Off”(APO) function
Procedure
1. Switch off the device.
2. Press and hold the “NULLbalance”key during the entire setting process and press the two
keys “R
PE
“ and “Auto” simultaneously.
The digital display shows “OFF” in the first line and the switch-off time (in minutes) in the
second line.
3. To increase the switch-off time, press the “V
ISO
“key (1...10min).
4. Release the “NULLbalance”key to save the settings.
5.3.2 Setting the date and the time
Procedure
1. Switch off the device.
2. Press and hold the “measured value memory”key and press the two keys “R
PE
“ and “Auto”
simultaneously.
The date and time format is displayed as follows:
MM.DD (month: 1...12, day: 1...31)
YYYY (year)
HH.mm (hours: 0...23, minutes: 0...59)
SS (seconds: 0...59)
3. Press the “R
PE
”key to select a date or time field.
As soon as the field is flashing, the value for that field can be set.
4. Press the “Save”key to increase the value or the “Measured value memory”key to
decrease the value.
Every change resets the field for the seconds to zero.
5. To save the entered values, switch off the device.