Carry out correction as follows:
1
*)( TC
MPMC
ϑ
ϑ
=
ρ
C
= corrected input value
ρ
M
= measured density value
ϑ
P
= product temperature during density determination
ϑ
M
= reference temperature
TC
1
= temperature coefficient
The display always relates to ρ
M
.
Example: Temperature coefficient = 0.5 · 10
-3
ρ
M
in
g/cm
3
ϑ
P
in °C
ϑ
M
in °C
ρ
C
in g/cm
3
1.30 75 70 1.2975
1.32 70 70 1.3200
1.35 65 70 1.3750
Since in this case correction has already been done manually, temperature compensation
must be turned off and then on again in the instrument configuration.