42
3.2 Push <sk2> to select Product Data and <more> to call parameters sequentially.
Select product
Select current output 0 –20 or 4 – 20 mA
Define current output limit values 0/4 and 20 mA
Define current output error message
Define Relay 2: Function (Min), switch point (e.g. 1.2 g/cm
3
)
Hysteresis (e.g. 5 %)
Define Relay 3: unction (Max), switch point (e.g. 1.6 g/cm
3
)
Hysteresis (e.g. 5 %)
Only for suspension measurements:
Enter Solid density in g/cm
3
for selected product
Enter liquid density in g/cm
3
Enable Water TC when solution is to be measured with water as carrier liquid.
Enter reference temperature
In all other cases:
Calculate and enter temperature coefficient TC1 (linear)
Calculate and enter temperature coefficient TC2 (square)
Enter reference temperature
Push <done> to return to menu group.
6.2.2 Calibration
One-point calibration can be carried out for preliminary adjustment of the measuring
ranges to the reading. Two- or multi-point calibrations have to be performed for accurate
and final adjustment. Different procedures are required, depending on the selected cali-
bration method. For more information on the calibration methods see chapter 8
CALIBRATION.
One-point calibration has to be performed if only one value pair is available or the value
pairs are close together. Enter an absorption coefficient as a1-value; for common prod-
ucts, this value may be taken from the table in the Appendix. The calculation formulas
for other products are listed in chapter 8 CALIBRATION.
Please read chapter 8 CALIBRATION in any case before performing calibration!
Prerequisite: Pipeline must be filled completely or level in container must be at least
200 mm above the measuring point.
Select menu Calibrate / Live Display with <more>.
Push <sk1> to call the Calibrate submenu.
Select product.
Specify if it is a suspension measurement.
Select unit.
Push <sk1> to select the Data input submenu.
Deselect Calibr. Data transfer with “No”
Read in 1. Rate: cursor appears in the row “1. Rate = cps”.
Push <run>. Count rate and temperature (if a temperature sensor is connected) are
read in. Wait until the measured value has become stable (20 to 50 s). While reading-in
the count rate, the density of the products must not change.
While reading in the count rate, take a sample of the product from the pipeline and de-
termine its density or concentration.