EasyManua.ls Logo

Binder KBF 720 (E5.2) - 19.4.2 KBF P technical data

Default Icon
103 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
KBF / KBF P (E5.2) 02/2012 page 93/103
19.4.2 KBF P technical data
Unit size
240
720
Exterior dimensions
Width
mm / inch
925 / 36.42
1250 / 49.21
Height (incl. castors)
mm / inch
1460 / 57.48
1925 / 75.79
Depth
mm / inch
800 / 31.50
887 / 34.92
Depth including door handle, I-triangle, connection
mm / inch
880 / 34.65
939 / 36.97
Wall clearance rear (minimum) (spacers)
mm / inch
100 / 3.94
100 / 3.94
Wall clearance side (minimum)
mm / inch
100 / 3.94
200 / 7.87
Steam space volume
l / cu.ft.
348 / 12.3
918 / 32.4
Quantity of doors
1
2
Quantity of inner glass doors
1
2
Interior dimensions
Width
mm / inch
650 / 25.60
973 / 38.31
Height
mm / inch
785 / 30.91
1250 / 49.21
Depth
mm / inch
485 / 19.09
576 / 22.68
Interior volume
l / cu.ft.
247 / 8.7
700 / 24.7
Quantity of racks (standard / max.)
2/7
3/12
Quantity of light cassettes
2
3
Load per rack
kg / lbs.
30 / 66
45 / 99
Permitted total load
kg / lbs.
100 / 220
150 / 331
Weight (empty)
kg / lbs.
213 / 470
374 / 825
Temperature data (without humidity)
Temperature range without light cassettes 1)
°C / °F
0 to +70 / 32 to 158
Temperature range with light cassettes, with illumination 1)
°C / °F
+10 to +60 / 50 to 140
Max. heat compensation up to 40 °C / 104 °F, with illumi-
nation
W
400
1000
Climatic data (with humidity)
Temperature range without light cassettes 1)
°C / °F
+10 to +70 /
50 to 158
+10 to +70 /
50 to 158
Temperature range with light cassettes, with illumination 1)
°C / °F
+10 to +60 /
50 to 140
+20 to +60 /
68 to 140
Temperature fluctuation
with illumination
at 25 °C / 77 °F and 60% r.H.
± K
0.2
0.2
at 40 °C / 104 °F and 75% r.H.
± K
0.2
0.2
Temperature uniformity
(variation) with illumination
at 25 °C / 77 °F and 60% r.H.
± K
0.6
1.2
at 40 °C / 104 °F and 75% r.H.
± K
0.6
1.2
Humidity range without light cassettes
% r.H.
10 to 80
10 to 80
Humidity range with light cassettes, with illumination
% r.H.
10 to 75
10 to 75
Humidity fluctuation with
illumination
at 25 °C / 77 °F and 60% r.H.
± % r.H.
1.5
≤ 2.0
at 40 °C / 104 °F and 75% r.H.
± % r.H.
2.0
≤ 2.0
Recovery time after doors
were open for 30 sec. 2)
at 25 °C / 77 °F and 60% r.H.
minutes
3
1
at 40 °C / 104 °F and 75% r.H.
minutes
4
5
Illumination data per light cassette
ICH compliant illumination for photo stability testing 3)
Lux
7500
9900
UVA W/m
2
1.4
2.1

Table of Contents

Related product manuals