10
Precautions for Electric Safety.
Please check the following points before use.
l Is the product supplied with the appropriate power? (100-240V~)
l Are all connecting parts (power cables or the product) properly
connected?
l Is the product completely grounded? (If not, noise may occur.)
l Before turning the power on, is the accessory for measuring the
target parameter well connected to the product?
To prevent electric noise during use, the product should be installed
apart from dynamo, X-ray equipment, broadcast equipment or portable
cables. An inaccurate result may occur when these equipment are placed
near the product. BPM-1200(1010) requires an independent power circuit
and a stable earth connection. Sharing power with other electric
equipment may result in inexact results.
BPM-1200(1010) is classified as follows :
◈ Type 1 for electric shock, CF grade for ECG, Resp, ICO and IBP
and BF grade for EtCO2, SpO2, NIBP and Temp.
◈ Please do not use this product near inflammable anesthetics
and diluents.
◈ The noise level is “B”Class according to the IEC/EN 60601-1
(Safety of Electric Medical Product), and the noise redemption
is “b”Level according to the IEC/EN 60601-1-2
(Electromagnetic Compatibility Requirements).
◈ Protection grade against water is “IPX 1”