l How to set the alarm
To change the alarm setting, touch the screen once to
wave or display. And touch the Bottom Alarm.
Alarm Range
Parameter
Setting range
HIGH
LOW
ECG [bpm]
50 ~ 300, OFF
OFF, 30 ~ 280
SpO2 [%]
21 ~ 99, OFF
OFF, 20 ~ 98
IBP
[mmHg]
Systolic
40 ~ 295, OFF
OFF, 30 ~ 290
Diastolic
30 ~ 289, OFF
OFF, 25 ~ 280
Mean
35 ~ 290, OFF
OFF, 26 ~ 285
RESP [rpm]
5 ~ 149, OFF
OFF, 1 ~ 145
NIBP
[mmHg]
Systolic
40 ~ 295, OFF
OFF, 30 ~ 290
Diastolic
30 ~ 285, OFF
OFF, 25 ~ 280
Mean
35 ~ 290, OFF
OFF, 30 ~ 285
TEMP [℃]
20.0℃ ~ 49.9℃, OFF
OFF, 10.0℃ ~ 40.0℃
EtCO2 [%]
2% ~ 99%, OFF
OFF, 1% ~ 98%
For the sake of the user’s convenience. The function of Alarm
On/Off functions are added to each parameter. The state of Alarm
On/Off does not affect the setting values of the alarm.