(2) Alarm
Different alarms are provided in BPM-1200(1010) for the sake of the
user’s convenience. Please read carefully the contents of this section in
order to learn how to interpret and react appropriately to the alarms.
◈ The alarms in the product conform to the international standard
“IEC 60601-1-8 (2006).
- A High/Medium/Low/Information alarm and Information alarm
were adopted in BPM-1200(1010).
◈ We recommend that you listen to the alarms before using the
product.
l High Alarm
- The High alarm will sound when the measurement values
exceed the proper range.
- In High alarm, 5 alerting sounds ring for 2 seconds, for a
total of 10 times. After 8seconds, the alerting sounds are
repeated.
l Information Alarm
- Information Alarm rings to notify you of any announcements
or if there has been an error measurement.
- The information alarm sound like a chime bell, and rings at
an interval 5seconds to alert the user.
- The information alarm rings for the following situations:
a. If there is a failure in the NIBP measurement or
overpressure.
b. At Lead fault or Lead off during measurement.
c. If the finger is not attached to the finger probe during
measurement of SpO2.
d. If the battery is low.
- When the information alarm rings, it can be cancelled by
operating the function switch.