69
1) Measurement Parameter
ECG, SpO2, NBP, 4IBP, RESP, 4TEMP, EtCO2, Multi Gas, ICO, HRV,
OXY-CRG, APG detect analysis, ST level display, mini trend, ECG
Recall, Drug Dose, Color change, Temp TD, NIBP Text trend.
2) Monitor Performances Specifications (Display Method)
- TFT Color LCD Display
- Waveform Display Method : Scrolling Method
- Display Size : 12.1inch (10.4inch)
- Resolution : VGA (800Х600)
- Sweep Speed : 3.12, 6.25, 12.5, 25, 50mm/sec
- Display waveform Time : 4 sec
- Waveform Freeze : Function of waveform freeze
- Waveform Display : ECG, SpO2, IBP 4ch, Resp, EtCO2
- Waveform Traces : 12 Trace
- Measurements Value Display : Heart Rate, SpO2, Pulse Rate, NIBP,
IBP, Temp, EtCO2, FiCO2, Multi Gas
3) ECG Measurement
- ECG Lead : 3 lead / 5 lead( I, II, III, / aVR, aVL, aVF, C1 )
10 lead (I, II, III, aVR, aVL, aVF, C1, C2, C3, C4, C5 C6)
- Input : Installed the circuit for protecting form the high frequency
electric potential of ESU, Defibrillator and so on.
- Frequency Filter : High frequency noise remove filter / Install the
50/60Hz power noise remove filter.
- Input Impedance : Greater 5 M ohm (50/60 Hz)
- Input for Initial Electric currency : Below 50 μA
- CMRR : 120dB
- Frequency Bandwidth) : 0.5 ~ 20 Hz, 0.5Hz ~ 40Hz, 0.05Hz ~
80Hz