68
Problems in IBP Measurement
Sign
How to teat
Blood pressure
measurement is
unstable.
Pour out the contents of the tube carefully and
remove the bubble, then shorten the tubing length.
Severe noise or bias in
the waveform.
Do not shake or touch the connection area of the
IBP sensor.
Problems in EtCO2 Measurement
Sign
How to treat
Check Adaptor
The filter-line and Airway Adaptor is not connected
to the main body. Try again.
Failure
Contact the agency.
Occlusion
Check if the sampling line and filter is obstructed,
and clean it up if possible. It should be changed if
necessary. To remove this message, disintegrate
the filter-line and then reconnect it.
Problems in PRINTER Measurement (Optional)
Sign
How to treat
No printing
1. Check if the message “PRINTER” is displayed
on the screen.
2. Check if the power LED lights up on the printer.
3. Check if the paper is set normally.
4. Check if the printer door is closed.
Noise during printing
1. Check if the roller is contaminated with foreign
materials or if the gear is out of position.
2. Check if the problem is due to a paper jam.
(change the paper)