Biotage ELSD-A120 - General Problems and Solutions

Biotage ELSD-A120
88 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Biotage
®
ELSD-A120 User Manual 47
Troubleshooting and Diagnostics
6
General Problems
General Problems
Baseline Noise
Baseline Spikes
Probable cause Suggested actions
1 Poor nebulization Increase the temperature of the
nebulizer by 10 °C until the baseline
noise decreases
2 Insufficient evaporation Increase the temperature of the
evaporator by 10 °C until the
baseline noise decreases
Increase the gas flow rate
Decrease the nebulization
temperature
3 Non-volatile additive in the mobile
phase
Use a volatile mobile phase
4 Pressure difference created inside
the nebulizer chamber
Ensure that the end of the liquid
waste tube is not immersed in
liquid
Ensure that the exhaust tube at
the rear of the detector is not
blocked, or extraction is too
strong
5 Pulsations, especially in
microbore applications where low
flow rates are used
Use a pulsation free nitrogen
supply
Probable cause Suggested actions
1 Particulate matter in the gas
supply
Filter the incoming gas, or change
the supply
2 Poor nebulization
Check that the inlet gas flow is
> 60 psi (4.1 bar) and stable
3 Insufficient evaporation Increase the temperature of the
evaporator by 10 °C until the
baseline noise decreases
4 Non-volatile additive in the mobile
phase
Replace the additive in the mobile
phase with a volatile buffer