ELSD-A120 User Manual 55
Error Information
7
Error Messages
Error Messages
Internal Temperature Exceeded Lower Limit
Error ID: 10
Air temperature inside the detector is ≤10 °C.
Internal Temperature Exceeded Upper Limit
Error ID: 11
Air temperature inside the detector is >40 °C.
On-board Vapor Sensor Failed
Error ID: 12
The vapor sensor located on the main control board has failed.
Probable cause Suggested actions
1 The environmental temperature is
outside the specified operating
limits of the detector
• Increase the ambient
temperature where the detector
is located
• Move the detector to a warmer
location
Probable cause Suggested actions
1 The environmental temperature is
outside the specified operating
limits of the detector
• Decrease the ambient
temperature where the detector
is located
• Move the detector to a cooler
location
Probable cause Suggested actions
1 Vapor sensor not connected to the
main board
Please contact your Biotage service
representative
2 Defective vapor sensor Please contact your Biotage service
representative