Biotage
®
ELSD-A120 User Manual 49
Troubleshooting and Diagnostics
6
General Problems
Large Baseline Offset
Peak Tailing
4 Insufficient smoothing The broader the peaks, the higher
the smoothing value is required.
Increase smoothing to 50 for flash
separations
5 Sample precipitation during
nebulization
Reduce the sample concentration
or inject the same loading in
larger volume
Check the sample solubility in the
mobile phase eluents
Probable cause Suggested actions
Probable cause Suggested actions
1 Inefficient evaporation Increase the evaporator
temperature and/or gas flow
2 High concentration of non-volatile
buffer or stabilizer
Use a lower concentration of the
stabilizer or unstabilized solvent or
use a more volatile buffer
3 Contaminated diffuser Perform cleaning procedure
4 Optics heater failed Please contact your Biotage service
representative
Probable cause Suggested actions
1 Eluent particles lingering in the
optical chamber
Increase the gas flow rate
2 Poor chromatography Optimize the flash separation