Troubleshooting and Diagnostics
6
General Problems
Large Baseline Offset
Peak Tailing
4 Insufficient smoothing The broader the peaks, the higher
the smoothing value is required.
Increase smoothing to 50 for flash
separations
5 Sample precipitation during
nebulization
• Reduce the sample concentration
or inject the same loading in
larger volume
• Check the sample solubility in the
mobile phase eluents
Probable cause Suggested actions
Probable cause Suggested actions
1 Inefficient evaporation Increase the evaporator
temperature and/or gas flow
2 High concentration of non-volatile
buffer or stabilizer
Use a lower concentration of the
stabilizer or unstabilized solvent or
use a more volatile buffer
3 Contaminated diffuser Perform cleaning procedure
4 Optics heater failed Please contact your Biotage service
representative
Probable cause Suggested actions
1 Eluent particles lingering in the
optical chamber
Increase the gas flow rate
2 Poor chromatography Optimize the flash separation