7 Error Information
Error Messages
Rear Vapor Sensor Failed
Error ID: 13
The vapor sensor located on the rear panel of the detector has failed.
Vapor Detected
Error ID: 14
Solvent vapor threshold exceeded inside the detector.
Leak Detected
Error ID: 15
A leak was detected inside the detector.
Probable cause Suggested actions
1 Vapor sensor not connected to the
main board
Please contact your Biotage service
representative
2 Defective vapor sensor Please contact your Biotage service
representative
Probable cause Suggested actions
1 External vapor being drawn into
the detector
Remove any source of solvent vapors
close to the detector
2 Solvent leak inside the detector Please contact your Biotage service
representative
3 Exhaust tube not fitted Fit the black exhaust tube
Probable cause Suggested actions
1 Loose nebulizer fittings Please contact your Biotage service
representative
2 Blocked nebulizer causing leak at
capillary fittings
Please contact your Biotage service
representative