Maintenance Operating instructions SC II
92 / 107 61429850113 en-US
or
Aerosol can, 200ml
Order no.: 9400 8900133
Illustration79: Aerosol can
Thermal strip cleaning set
consisting of:
5 cleaning buds made of wood with leather
coating
1 pcs. IPA cleaning solvent 12 ml
Order no.: 9400 8919 000
Illustration80: Thermal strip cleaning set
Moisten the tip of the cleaning bud with a
few drops of the cleaning solvent.
Clean the thermal strip immediately with
cleaning bud.
Illustration81: Application
8.2 Maintenance
The load plate must be removed every 4 weeks and loose dirt removed with a brush or vac-
uum cleaner.
Replace the load plate and check the zero point of the scale. Switch the scale off and on if
necessary.
8.2.1 Testing the weight display
Place reference weight of 2x3 kg on the scale and read off the display at 3 and 6kg.
This test must be repeated at regular intervals.