101
To recall setup data from USB flash drive, follow the steps
below:
1. Press the “SAVE/RECALL” button.
2. Press the “Type” button to select “Setups”.
3. Insert USB flash drive to USB host port of the
oscilloscope and wait that the oscilloscope has initialized
USB flash drive (about five seconds).
4. Press the “Save to” option button to select “File”.
5. Press the “Save” option button then you’ll go into the
Save/Recall interface.
6. Choose the file you want then press the “Load” option
button (within 5 seconds there will be a message “Read
data success”). Now the setup data have been recalled
from the USB flash drive.
Recall Factory
You can use this option to recall the factory settings
Table 3.45 Factory Function Menu
Option
Setting
Description
Type
Factory
To view the Factory setup.
Depth Recover
Clears all memory, including
setups, waveforms and masks.
Load
Recall the Factory setup.