SCPI Command Reference
67
MEASure command
Command syntax:
:MEASure:<function>?
Command parameter :
<function> = VOLTage:DC DC voltage
VOLTage:AC AC voltage
CURRent:DC DC current
CURRent:AC AC current
RESistance 2-wire resistor
FRESistance 4-wire resistor
FREQuency Frequency
PERiod Period
DIODe Diode testing
CONTinuity Continuity test
Return String:
Returns a measured reading.
Description:
This command combines all of the other signal oriented measurement commands to perform a “one
time” measurement and acquire the reading.
When this command is sent, the following commands execute in order:
:ABORt
:CONFigure:<function>
:READ?
When :ABORt is executed, if continuous initiation is disabled, the instrument goes into the idle
state. If continuous initiation is enabled, the operation re-starts at the beginning of the Trigger
Model.
When :CONFigure is executed, the instrument goes the selected measurement mode.
When :READ? is executed, another :ABORt is performed, then an :INITiate, and finally a FETCh? to
acquire the reading.
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the BK Precision 5492B and is the answer not in the manual?

BK Precision 5492B Specifications

General IconGeneral
BrandBK Precision
Model5492B
CategoryMultimeter
LanguageEnglish

Related product manuals