RS-232 Settings
In order for the computer to communicate with the power supply, both must be set to the same
RS-232 settings. These communication settings are:
- Baud rate must be one of 4800, 9600, or 38400
- 8 data bits
- One stop bit
- No parity
Refer to the Baud Rate section in the Menu for instructions on how to select a different baud
rate.
Note: The remote interface does not have a flow control mechanism. The programmer
should be aware of this limitation and allow sufficient command process time. If the
remote commands are sent too fast the power supply’s internal buffer may overrun and
cause a communication error. Therefore, adding a delay between commands to allow for
sufficient time to process commands is required.