3
Operations & Maintenance Manual: Important Information
CAUTIONS & WARNINGS
Warnings indicate that injury, death, or other serious adverse reactions may result from use or misuse of
the device. Cautions indicate that use or misuse of the device may cause a problem, such as a malfunction,
failure, or damage to the product. Throughout the manual, pay attention to sections labeled Important, as
these contain reminders or summaries of the following cautions as they apply to a specific component or
usesituation. Please heed the following warnings and cautions.
PRECAUTIONS
Potential Device Interference. Bluetooth
®
and wireless LAN devices operate within the same
radio frequency range and may interfere with one another.
If you are using the BladderScan BVI9400 Bluetooth link and wireless LAN devices
simultaneously, you may experience less‑than‑optimal network performance or even lose your
network connection. If this happens, you may need to move the BladderScan and Scan Point
®
host computer to an area away from the 2.4‑GHz wireless LAN devices (40m [44yd], ormore).
CAUTION
Use of the following cleaning methods or solutions may cause device damage not covered by
the BladderScan BVI9400 warranty.
Do not immerse the instrument in disinfectant solution.
Do not use Cidex Plus
®
to disinfect the instrument. Cidex Plus will damage the plastic enclosure.
Do not subject any part of the instrument to steam sterilization or ethylene oxide sterilization.
CAUTION
When using the BladderScan BVI9400 with optional Scan Point
®
software, your computer must
be minimally certified to IEC 609501 or 610101 standards. This configuration ensures that
compliance to the IEC 606011 system standard is maintained. Anyone connecting additional
equipment to the BladderScan BVI9400 signal input port or signal output port configures
a medical system, and is therefore responsible for ensuring that the system complies with
IEC606011. If you need assistance, contact your biomedical staff, Verathon representative, or
Verathon Customer Care.
CAUTION
The BladderScan BVI9400 and related devices may contain mineral oils, batteries, and other
environmentally hazardous materials. When the instrument or accessories have reached the end
of their useful service life, see the section Device Disposal on page61.
CAUTION