Operations & Maintenance Manual: Product Specifications
ELECTROMAGNETIC IMMUNITY
Table 30. Guidance and Manufacturer’s Declaration Electromagnetic Immunity
The system is intended for use in the electromagnetic environment specified below. The customer or the user
of the system should ensure that it is used in such an environment.
IMMUNITY TESTS IEC 60601 TEST LEVEL
COMPLIANCE
LEVEL
ELECTROMAGNETIC ENVIRONMENT –
GUIDANCE
Electrostatic discharge
(ESD)
IEC 61000‑42
± 6kV contact
± 8kV air
In compliance
Floors should be wood, concrete,
or ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.
Electrical fast
transient/burst
IEC 61000‑4‑4
± 2kV for power supply
lines
± 1kV for input/output
lines
In compliance
Mains power quality should be that
of a typical commercial or hospital
environment.
Surge
IEC 61000‑45
± 1kV line(s) to line(s)
± 2kV line(s) to earth
In compliance
Mains power quality should be that
of a typical commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000‑411
<5% Ut (>95% dip in Ut)
for 0.5cycle
40% Ut (60% dip in Ut)
for 5cycles
70% Ut (30% dip in Ut)
for 25cycles
<5% Ut (>95% dip in Ut)
for 5s
In compliance
Mains power quality should be
that of a typical commercial or
hospital environment. If the user of
the BladderScan BVI9400 system
requires continued operation during
power mains interruptions, it is
recommended that the BladderScan
BVI9400 system be powered from
an uninterruptible power supply or a
battery.
Power frequency
(50/60Hz) magnetic
field
IEC 61000‑4‑8
3A/m In compliance
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital environment.
Conducted RF
IEC610004‑6
3Vrms
150kHz to 80MHz
3V
Portable and mobile RF
communications equipment
should be used no closer to any
part of the BladderScan BVI9400
system, including cables, than the
recommended separation distance
calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation
distance d (m)
d=1.2P