EasyManua.ls Logo

Bosch EPS 205 - Decommissioning; 7.1 Change of location; 7.2 Disposal and scrapping; Technical data

Bosch EPS 205
636 pages
Go to English
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1 689 989 185 2018-03-24| Robert Bosch GmbH
Decommissioning | EPS 205 | 69 en
7. Decommissioning
7.1 Change of location
Heed the following if the EPS 205 is to be relocated:
R Unplug the electrical connection.
R Disconnect the pneumatic connection.
R Drain the test oil out of the test oil tank (see sec.
6.2.6 , steps 1 to 8).
R Re-fit the transportation safeguards for the drive
unit (see sec. 4.4. ).
R The EPS 205 is only to be transported in a crate.
If the EPS 205 is to be sold or passed on, all the
documentation included in the scope of delivery must
be handed over together with the unit.
7.2 Disposal and scrapping
Before scrapping, the test oil in the EPS 205 must be
drained off. The pipe filter at the flowmeter, the intake
filter in the test oil tank and the filter mats for external
extraction in the jet chamber unit must also be removed
and disposed of.
i Test oil as per ISO 4113 is classified under used oil
iicollection category 1. In category 1, the used oil
must not contain any foreign substances such as
used oils of a different category, gasoline or diesel
fuel. The corresponding refuse code number can be
taken from the safety data sheet of the test oil.
8. Technical data
8.1 EPS 205
Function Specification
Rated three-phase AC voltage 380 VAC – 460 VAC
Rated current 15 A
Fuse 16 A
Number of phases 3P - PE
Input frequency 50 Hz / 60 Hz
Rated power 4,2 KW
Compressed air 0.5 MPa − 0.8 MPa
Storage temperature -25 °C – 60 °C
Operating temperature 5 °C – 40 °C
*)
Ambient temperature
for measurement accuracy
10 °C – 35 °C
Max. perm. relative humidity ≤90 % (25 °C,
24hours duration)
Degree of protection IP 22
Oil pressure 180 MPa
Maximum speed 3500 min
-1
Control voltage 24 VDC
Atmospheric pressure
corresponding to an altitude of
700 hPa – 1060 hPa
≤2200 m
*)
Longer conditioning times must be expected at ambient
temperatures above 25 °C.
Function Specification
Tightening torques for high-pressure
connection, screw plug and hoses
25 Nm − 30 Nm
Test oil tank capacity 6.5 l
8.2 Noise emissions
Function Specification
Emission sound pressure level
at workplace as per DIN EN ISO 11201
< 71.5 dB(A)
Sound power level as per DIN EN ISO 3744 < 84.1 dB(A)
8.3 Dimensions and weights
Function Specification
EPS 205 (H x W x D) 580 x 560 x 780 mm
Weight of EPS 205 incl. packaging
Weight of EPS 205
(without test oil and packaging)
210 kg
138 kg
Oil collector (H x W x D) 30 x 570 x 720 mm
EPS 205, accessories and packaging should
be sorted for environmental-friendly recycling.
Do not dispose EPS 205 into household
waste.
Only for EC countries:
The EPS 205 is subject to the European
directive 2012/19/EC (WEEE).
Dispose of used electrical and electronic
devices, including cables, accessories and
batteries, separately from household waste.
Make use of the local return and collection
systems for disposal.
Proper disposal of EPS 205 prevents
environmental pollution and possible health
hazards.

Table of Contents