EasyManuals Logo

Braun Infusomat Space Service Manual

Braun Infusomat Space
136 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #7 background imageLoading...
Page #7 background image
Important Preliminary Remarks
0
Infusomat® Space 6.0 0 - 7EN
Each workstation must be equipped according to the recommen-
dations with the necessary static protective measures, if
ESD
components or boards are handled.
Each workstation must be equipped with a conductive table sur-
face. The conductive surface, the solderi
ng iron or the soldering
stations must be grounded via protective resistors.
Chairs must be of antist
atic design. The floor or floor mats should
be of electrically conductive material.
Personnel must wear conductive wristbands which are
connected
to a central ground potential via protective resistors, e.g. the
ground contact of a wall outlet. Furthermore it is recommended
that personnel wear cotton clothing and electrically conductive
shoes to prevent electrostatic charge.
Spare Parts and Test Equipment Only use original spare parts from the
manufacturer. Do not
tamper with assembly groups which can only be exchanged com-
pletely. The spare parts required a
re listed in the repair descrip-
tions.
When assembling spare parts with a serial or
batch number (see
label on the packaging or number plate) the serial number of the
affected medical device has to be documented.
Service personnel are responsible for
the calibration of their test
equipment. Original test equipment can be calibrated at the
works of B. Braun. Further information is available upon request.
Safety data sheets You can request the material safety data sheets for the consum-
ables described in this Service Manua
l free of charge from
B. Braun. You can download the latest
version of each document
as a PDF in the service portal under
https://extranet.bbraun.com.
Setting Off Additional notes and warnings are set off as follows:
Note
Is used for additional or special notes concerning information and
working steps.
CAUTION
Is used for working steps which may result in damage to the unit,
system or to a connected device.
WARNING
!
Is used for working steps which may result in personal injury.
Fig.: 0 - 1

Other manuals for Braun Infusomat Space

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Braun Infusomat Space and is the answer not in the manual?

Braun Infusomat Space Specifications

General IconGeneral
TypeInfusion Pump
Power SupplyAC/DC adapter, rechargeable battery
DisplayColor graphic display
AlarmsAudible and visual alarms
Flow Rate Accuracy±5%
Occlusion PressureAdjustable
VTBI Range0.1 – 9999 ml
KVO Rate0.1 – 5 ml/h (adjustable)
BatteryLithium-ion
Operating ModesContinuous
CertificationsCE, ISO 13485, IEC 60601-1

Related product manuals