Disassembly / Assembly
3
3 - 12 Infusomat® Space 6.0
EN
3.6 Battery Module
Designation Order No.
Battery compartment cover ISP, cpl. . . . . . . . . . . . . . . 3452 1321
(incl. emergency release crank)
Battery module SP (NIMH), without fixing pin . . . . . . 3452 0856
Battery module SP (NIMH)
, with fixing pin
*
. . . . 3452 0856A
Battery module SP (Li-ion)
with WLAN and fixing pin . . . . . . . . . . . . . . . . . . . 8713182A
Users are informed as follows: If a Infusomat® Space pump run-
ning from battery only and showing
full battery charge stops
while the operating unit is closing and triggers an alarm, the user
must remove the pump from use and return it to the Medical En-
gineering Department for replacement of the battery. In this case
the battery must be replaced regardless of its age or any further
tests.
Disassembly
1. Press the lock (Fig.: 3 - 12 / Item 2) on the battery compart-
ment cover (Fig.: 3 - 12 / Item 1) with a pointed tool and re-
move the battery compartment cover.
Fig.: 3 - 12
Legend of fig. 3 - 12:
Item Designation
1 Battery compartment cover
2 Lock for battery compartment cover
2
1
*
The new battery module, Article number 34520856A, is not downwards compat-
ible. Please only use for devices with a new battery contact strip.
Further information is available on request.

Other manuals for Braun Infusomat Space