Infusomat® Space 6.0 2 - 1EN
2 - Unit Diagnosis / Calibration
Alarms and Error Codes
The alarms of the Infusomat® Space are classified in 5 categories.
These categories are listed hereafter according to their impor-
tance.
- Alarm advice
In case of unacceptable inputs corresponding messages are
d
isplayed (e.g. “Caution! Rate out of range“, “The parameter
cannot be changed“) and a beep sounds.
-Pre-alarm
Pre-alarms are triggered several minutes (depending on
the
service settings) before the operating alarms.
- Reminder alarm
A reminder alarm is triggered if
the device is not operated for
two minutes when input or operation was not finished.
- Operating alarm
In case of an operating alarm the infusion
is stopped. An au-
dible signal is released, the red LED flashes and a staff call is
trig
gered. The message “Alarm” and the cause of the alarm
appear on the display.
- Device alarm
The most important alarms and error codes
as well as their mean-
ing and possible fault clearance are specified in the following
lists.
2 - 16

Other manuals for Braun Infusomat Space

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Braun Infusomat Space and is the answer not in the manual?

Braun Infusomat Space Specifications

General IconGeneral
TypeInfusion Pump
Power SupplyAC/DC adapter, rechargeable battery
DisplayColor graphic display
AlarmsAudible and visual alarms
Flow Rate Accuracy±5%
Occlusion PressureAdjustable
VTBI Range0.1 – 9999 ml
KVO Rate0.1 – 5 ml/h (adjustable)
BatteryLithium-ion
Operating ModesContinuous
CertificationsCE, ISO 13485, IEC 60601-1

Related product manuals