Unit Diagnosis / Calibration
2
2 - 10 Infusomat® Space 6.0
EN
The Most Important Error Modes
The following list specifies the most important error modes and
their clearance.
Calibration
Test Equipment
Designation
Slide clamp
(of Infusomat® Space Line)
Starting Calibration
Calibration must be carried out with power supply connected,
since the calibration can be interrupted suddenly if the unit is
battery operated and the battery becomes discharged so that the
device is switched off.
Error/Fault Possible Cause Fault Clearance
1 The battery module discharges too fast The device has not been used for some
tim
e. The battery module has not been
discharged and charged regularly.
Discharge and cha
rge battery module
several times.
Replace b
attery module
2 Sudden loss in capacity despite fully
c
harged battery module
Loss of contact between battery module
and battery contact strip
Retrofit battery fastening (see „Ret-
rofitting of the battery fastening“
pg. 3 - 16)
Fo
r unit software version J0300003,
contact service hotline and perform a
software update, when appropriate
3 Message "Battery does not exist" Loss of contact between battery module
and b
attery contact strip
Retrofit battery fastening (see „Ret-
rofitting of the battery fastening“
pg. 3 - 16)
4 Noises while opening or closing the oper-
ating unit
Defective toothed wheels in lock bolt
driv
e
Replace lock bolt drive
5 Device alarm 2027 Air sensor defective Exchange air
sensor
6 Device alarms 2021 ... 2022 Upstream pressure sensor defective Replace upstream pressure sensor
7 Device alarms 2023 ... 2024 Downstream pressure sensor defective Replace d
ownstream pressure sensor
8 Device alarm 2111 Loss of contact between battery module
and b
attery contact strip
Retrofit battery fastening (see „Ret-
rofitting of the battery fastening“
pg. 3 - 16)
9 Device alarm 2119 Manufacturing tolerance of LC display
too larg
e
Replace LC display
Table 2 - 6

Other manuals for Braun Infusomat Space

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Braun Infusomat Space and is the answer not in the manual?

Braun Infusomat Space Specifications

General IconGeneral
TypeInfusion Pump
Power SupplyAC/DC adapter, rechargeable battery
DisplayColor graphic display
AlarmsAudible and visual alarms
Flow Rate Accuracy±5%
Occlusion PressureAdjustable
VTBI Range0.1 – 9999 ml
KVO Rate0.1 – 5 ml/h (adjustable)
BatteryLithium-ion
Operating ModesContinuous
CertificationsCE, ISO 13485, IEC 60601-1

Related product manuals