Unit Diagnosis / Calibration
Infusomat® Space 6.0 2 - 9EN
c) If the fault does not occur with the first two pumps, test
the other pumps affected by the communication problem
on the same slot until you have found the defective
pump.
Eliminating the CAN-Dead state
1. Connect the HiBaSeD in order to elim
inate the CAN-Dead
state.
2. Carry out calibration.
3. Take particular care to enter the serial number correctly
.
Entering the states in the history
In order to provide diagnostic information to service technicians,
the vario
us CAN states are entered in the history.
They can then recognise, for example, whether there was a dupli-
cate INR in the system or whether address assignment
function of
a Station is defective.
The following history entry has to be evaluated for this purpose:
History Entry Value Meaning Measure
CAN Error Code 1 Duplicate serial number in the system Calibrate the device using HiBaSeD and check
the serial numb
er in the process.
CAN Error Code 2 Duplicate address assigned in the sys-
tem
Check the SpaceStation
Address assignment defective
Table 2 - 5