Disassembly / Assembly
3
Infusomat® Space 6.0 3 - 19EN
3.7 Housing Upper Part
Designation Order No.
Housing upper part ISP . . . . . . . . . . . . . . . . . . . . . . . . . 3452 1313
Screws and cover caps
(see „Service Parts and Screw Kit“ pg. 3 - 7)
Note
Please pay attention to the corresponding notes during assembly
and installation (see „Housing Upper Part“ pg. 3 - 46).
Disassembly
1. Pierce 6 cover caps (Fig
.: 3 - 18 / Item 1) with a small screw-
driver and prise out the cover caps.
2. Unscrew 6 screws (Fig.: 3 - 18 / Item 2).
Note
The other 4 cover caps on the housing bottom side must not be
removed during this disassembly state.
Fig.: 3 - 18
Legend of fig. 3 - 18:
Item Designation
1 Cover cap
2 Screw DELTA PT 30x12 WN 5452 A2 TORX plus 10IP
1