Disassembly / Assembly
3
3 - 18 Infusomat® Space 6.0
EN
6. If the retrofitting kit is used, replace the battery slide on the
battery module (nominal dimension after replacement: 2.15
mm ±0.15 mm).
7. Clean the bonding areas with isopropanol and let them air.
8. Position the removal aid (Fig.: 3 - 17 / Item 1) over the fixa-
tion (Fig.: 3 - 17 / Item 2) and move it.
9. Press the removal aid onto the battery bottom in order to
avoid
bubbling during bonding.
10. After retrofitting, reinsert the battery module and place on
the
battery compartment cover.
11. Carry out a function test (in battery operation).
- Set the max. pump rate (duration of pump operation ap-
p
rox. 1 min). During pump operation simulate a vibratory
load on the pump (shaking).
Fig.: 3 - 17
Legend of fig. 3 - 17:
Item Designation
1 Removal aid
2 Fixation
1
2