Disassembly / Assembly
Infusomat® Space 6.0 3 - 17EN
2. In order to retrofit the top housing section, clean the bonding
areas (Fig.: 3 - 15 / Item 1) with isopropanol and let them air.
3. Position the bonding segments i
n the battery compartment
using a scalpel.
4. Bond in the segments (5x) in the positions shown.
5. After bonding, press on the segments in order to avoid bub-
bling.
Fig.: 3 - 15
Legend of fig. 3 - 15:
Item Designation
1 Bonding areas
Fig.: 3 - 16
Legend of fig. 3 - 16:
Item Designation
1 Position of top segment
2 Position of side wall segments
3 Position of bottom segments