Infusomat® Space 6.0 3 - 1EN
3 - Disassembly / Assembly
3.1 General
Remarks on Disassembly / Assembly
Before each disassembly and assembly of a unit subsystem chec
k
the connectors, plug contacts and connections for corrosion and
tight fit.
The necessary steps to disassemble the complete unit, all its sub-
systems and spare parts are detailed in the
following description.
Steps that are not necessary can be skipped.
Before disassembling the unit make sure that no Infusomat®
Space
Line is inserted and that the operating unit is in the open
position.
CAUTION
Always remove the battery module before opening the device in
order to avoid a short-circuit.
Note
Plastic thread-forming screws are used in this device. These are
special screws for plastic housings. Pay attention to the corre
-
sponding notes when you fit the screws.
3 - 58

Other manuals for Braun Infusomat Space

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Braun Infusomat Space and is the answer not in the manual?

Braun Infusomat Space Specifications

General IconGeneral
TypeInfusion Pump
Power SupplyAC/DC adapter, rechargeable battery
DisplayColor graphic display
AlarmsAudible and visual alarms
Flow Rate Accuracy±5%
Occlusion PressureAdjustable
VTBI Range0.1 – 9999 ml
KVO Rate0.1 – 5 ml/h (adjustable)
BatteryLithium-ion
Operating ModesContinuous
CertificationsCE, ISO 13485, IEC 60601-1