Disassembly / Assembly
Infusomat® Space 6.0 3 - 23EN
3.9 Inner Frame and Housing Front Panel
Designation Order No.
Screws
(see „Service Parts and Screw Kit“ ➨ pg. 3 - 7)
Please pay attention to the corresponding notes during assembly
and installation
(see „Inner Frame and Housing Front Panel“ ➨
pg. 3 - 44).
Disassembly
When dismounting the inner frame and the housing front panel
please pay attention to the length of the connection cables to the
processor PCB.
1. Fold the locking of the ribbon cable connector (Fig.: 3 - 22 /
Item 4) to the air inline sensor up and pull the ribbon cable
off the connector.
2. Pull the safety clamp connector (Fig.: 3 - 22 / Item 5) off the
processor PCB.
3. Pull the pump drive connector (Fig.: 3 - 22 / Item 6) off the
processor PCB.
Fig.: 3 - 22
Legend of fig. 3 - 22:
Item Designation
1 Connector for ribbon cable, operating unit
2 Lock bolt drive connector
3 Pressure sensor connector, downstream
4 Air inline sensor connector
5 Safety clamp connector
6 Pump drive connector
7 Pressure sensor connector, upstream