Disassembly / Assembly
3
3 - 24 Infusomat® Space 6.0
EN
4. Loosen four screws (Fig.: 3 - 23 / Item 2)and lift the inner
frame together with the housing front panel
from the bottom
inner frame(Fig.: 3 - 23 / Item 6).
5. Unscrew one screw (Fig.: 3 - 23 / Item 5) and remove the PCB
of the direction of rotation sensor for the pump drive from
th
e pump frame.
The PCB of the direction of rotation sensor is connected to the
processor PCB via the ribbon cable.
6. Pull the lock bolt drive connector (Fig.: 3 - 22 / Item 2) off the
processor PCB.
7. Pull the pressure sensor connector
, upstream (Fig.: 3 - 22 /
Item 7) off the processor PCB.
8. Pull the pressure sensor connector, downstream (Fig.: 3 - 22
/ Item 3) off the processor PCB.
Disassembly
1. Carefully remove the inner frame (Fig.: 3 - 23 / Item 3) from
the housing front (Fig.: 3 - 23 / Item 1) panel.
The membrane in the housing front panel must not be used again.
Always use a new membrane for assembly.
Fig.: 3 - 23
Legend of fig. 3 - 23:
Item Designation
1 Housing front panel, cpl.
2 Fillister head screw M3x6 TORX
3 Inner frame, cpl.
4 PCB of direction of rotation sensor
5 Screw DELTA PT 25x7 WN 5451 TORX plus 8IP
6 Bottom inner frame
1
2
5
4
3
6
2