Page #16 background imageLoading...
Page #16 background image
Software
2
2- 2 Perfusor® compact, 2.1 gb
Error Messages and Alarms
In case of a unit malfunction a continuous signal is activated, and
the function processor displays an alarm and an error code. The
error code of the control microprocessor can be queried with the
F button. Please state both error codes if you have any questions.
Acknowledge alarm and switch device off.
Device Alarms of the Function Processor
LC display Description
1 Different syringe recognition
2 Different FP- and CMP condition
3 Rate of FP- and CMP different
4 Different function mode
5 Different rate of delivery
6 Different target volume
7 Different step volume (low)
8 Different motor steps
19 State/motor state
20 Invalid normal state
21 return from PlcMain
22 Unexpected reset
28 No sync at Plc_Down
29 No sync at Plc_On
30 Different CMP/FP mode ports
31 Invalid mode ports
32 Invalid variable values
33 Error in ROM test
34 Different software version
40 Unexpected interrupt
49 Faulty sensor sync
51 Motor on during reverse run
52 Step cumulation > 10 steps
53 Illegal setting of Mot_Ok
54 Different recognition of direction of rotation
55 Reverse polarity of motor
Table 2 - 1 (Part 1 of 2)
2.1

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Braun perfusor compact and is the answer not in the manual?

Braun perfusor compact Specifications

General IconGeneral
ModelPerfusor compact
Occlusion PressureUp to 1200 mmHg
AlarmsOcclusion
Syringe CompatibilityVarious sizes from 5 ml to 50 ml
ManufacturerB. Braun